APSL803Fully Automatic Pressure Calibration Device
Product Description
APSL803Fully automatic pressure calibration device, capable of communicating with any pressure controller, can fully automatically calibrate 4 pressure transmitters, fully automatically perform high static pressure tests on differential pressure transmitters, can fully automatically calibrate digital pressure gauges, automatically recognize readings from digital pressure gauges, can connect to a computer to automatically print calibration original records and calibration reports, this device can record 10,000 records, and export EXCEL files via USB. Uses a wireless barcode scanner to scan barcodes on samples, automatically retrieves sample information, and performs fully automatic calibration.
Except for manual intervention required for installing and removing the tested instruments, all other processes are fully automated without the need for personnel supervision, greatly improving detection efficiency. Suitable for use in metrology institutes.
Specifications
- Reference standard
- Auto-verifies pressure transmitters, digital pressure gauges & indicating gauges; per Chinese pressure-instrument verification regulations
- Pressure range
- 1 kPa ~ 70 MPa (model dependent)
- Accuracy
- Pressure Class 0.01 / 0.02 (controller dependent); current Class 0.02; voltage Class 0.02; 6-digit resolution
- Medium
- Air or oil (determined by the controller)
- Measure / output signals
- 4× current inputs (0-22) mA; 4× voltage inputs (0-5.5) V
- Control performance
- Transmitter points 0/25/50/75/100%; digital gauge 0-100% in 10% steps; auto-return & freely configurable; dwell 10-240 s
- Pressure units
- 14 units: kPa, psi, inHg, inH₂O, mmHg, mmH₂O, MPa, bar, mbar, atm, kg/cm², hPa, Pa, Torr
- Communication
- Built-in HART (replaces HART communicator); wireless barcode scanner; stores 10000 records; USB export to Excel; connects to TMMS-2000
- Protection
- Over-pressure (alarm at 110% of range), over-current (>20.5 mA), leakage & sensor-fault protection; stops pressurizing on fault
- Process connection
- M20×1.5 female quick connector; φ6 standard banana jacks
- Power output
- DC 24 V/30 mA loop supply; will not damage a standard gauge if mis-connected
- Dimensions
- 185 (H) × 275 (D) × 355 (W) mm
- Weight
- 3 kg
Product Details
Product Overview
APSL803The fully automatic pressure calibration device can communicate with any pressure controller, automatically calibrate 4 pressure transmitters, automatically perform high static pressure tests on differential pressure transmitters, automatically calibrate digital pressure gauges, automatically recognize readings from digital pressure gauges, connect to a computer to automatically print calibration original records and calibration reports. This device can record 10,000 records and export EXCEL files via USB. It uses a wireless barcode scanner to scan barcodes on samples, automatically retrieves sample information, and performs fully automatic calibration. Except for manual intervention required for installing and removing the tested instruments, all other processes are fully automated without the need for personnel supervision, greatly improving detection efficiency. Suitable for use in metrology institutes.
This device is designed entirely according to Chinese pressure gauge calibration procedures. During the upward stroke, it approaches the calibration point from bottom to top; during the downward stroke, it approaches the calibration point from top to bottom.
This device is compatible with TMMS2000 Used together with automatic calibration software, it forms a fully automatic pressure gauge verification system, realizing true automatic verification of pressure gauges. As long as the range is selected, pressing the "Auto" key can achieve automatic pressurization, depressurization, pressure control, data recording, error analysis, printing verification reports and certificates, etc., that is, it has a "one-key" pressure gauge verification function.
Overpressure protection is set: when the transmitter current exceeds 20.5mA , pressurization stops.




Technical features
A truly fully automatic pressure calibration station. Just press the auto key to automatically complete all tasks such as pressurization, depressurization, pressure control, data recording, error analysis, printing calibration reports and certificates.
Fully complies with Chinese pressure gauge calibration regulations.
Can calibrate pressure transmitters, digital pressure gauges, and pressure indicating gauges.
Built-in HART function, can fully replace HART communicator.
Has calibration result recording function, and can export results in EXCEL format from USB drive. Can manually set pressure setpoints, automatic pressurization can manually select range, automatically pressurize for upstroke and downstroke according to selected range's calibration points. Can preset calibration points for multiple pressure ranges, achieving fully automatic calibration.
Has complete protection functions, overpressure protection (if exceeding range's 110% an alarm will be issued), overcurrent protection (when the transmitter current exceeds20.5mA), leakage, sensor fault and other alarms. When overpressure, overcurrent, or sensor fault occurs, pressurization stops.
It can be connected to the TMMS-2000 automatic calibration system to form a fully automatic pressure calibration system. Calibration personnel can complete pressure calibration work with just a mouse.

Functional specifications
Working medium: air or oil, determined by the controller.
Pressure range:1 kPa~70MPa(Determined by model).
Display resolution: 6 digits.
Accuracy class: Pressure: class 0.01 or class 0.02, determined by the controller; Current: class 0.02, voltage: class 0.02.
Measurement units: 14 types: kPa, psi, inHg,inH2O、mmHg、mmH2O、MPa、bar、mbar、atm、kg/cm2、hPa、Pa、Torr。
Standard calibration points: for transmitters, it is 0%、25%、50%、75%、100% of the range, with automatic return. Can be set arbitrarily. For precision digital pressure gauges, it is 0%,10%,20%,30%,40%,50%,60%,70%,80%,90%,100%, automatic return. Can be set arbitrarily.
Custom calibration points: arbitrary.
Current input: 4 channels, range: (0~22) mA.
Voltage input: 4 channels, range: (0~5.5) V.
Safety protection: overvoltage, overcurrent, sensor fault and other protections.
HART function: can fully replace HART handheld communicator.
Automatic verification software: optional.
Automatic pressure control time at verification points: (10 to 240) seconds, adjustable.
Loop power supply:24VDC/30mA, if mistakenly connected to a standard meter, it will not damage the standard meter.
Pressure interface:M20*1.5 internal thread quick coupling.
Current/Voltage Interface: ф6 standard banana plug.
Weight: 3kg; Dimensions: 185 (H) * 275 (D) * 355 (W) mm.