APSL701High Static Pressure Differential Fully Automatic Pressure Calibration Station
APSL Fully Automatic Pressure Calibration BenchesAPSL701

APSL701High Static Pressure Differential Fully Automatic Pressure Calibration Station

Product Description

1. Product Overview The APSL series 701 high static pressure differential pressure fully automatic pressure calibration system is a patented product of our company.

The related patents include: 1) A pressure calibration bench with built-in pressure modules, patent number: 201220302065.8, 2) A fully automatic pressure calibration bench for high static pressure differential pressure transmitters, patent number: 201220302053.5, 3) A fully automatic pressure calibration bench based on HART communication, patent number: 201220302062.4, 4) A fully automatic pressure calibration bench, patent number: 201220327679.1, 5) The appearance of the fully automatic pressure calibration bench, patent number: 201230276258.6. This calibration bench has built-in high-accuracy static pressure sensors and differential pressure sensors.

When calibrating a high static pressure differential pressure transmitter, set the static pressure setpoint and the differential pressure range, then press the "Start" button, and the system will automatically perform the calibration: first open the bypass valve of the differential pressure, automatically apply static pressure, and when the static pressure reaches the set value, it will automatically calibrate according to the set number of calibration points, from 0%~100% of the differential pressure range, and automatically record data. If connected to a computer, it can automatically print the calibration report.

Specifications

Reference standard
Auto-verifies high static pressure differential pressure transmitters; 5 patents
Pressure range
Static pressure module (0-20) MPa; differential pressure module (0-2000) kPa
Accuracy
Pressure Class 0.02 / 0.05; current Class 0.02 / 0.05; 5-digit resolution
Measure / output signals
1 current input (0-22) mA; stability ±5 digits on touchscreen
Control performance
Standard points 0/25/50/75/100% auto-return; dwell 10-240 s; 9-point run ≈7 min
Pressure units
Differential: kPa; static: kPa or MPa
Protection
Over-pressure, over-current & pressure-module fault protection; gas-liquid separator tolerates water/oil in DUT
Process connection
M20×1.5 female quick connector; φ4 standard banana jacks; φ6 mm drain port
Power output
220 V/50 Hz; DC 24 V/30 mA loop supply; media: air or transformer oil
Operating temperature
-5 ~ 50 °C
Dimensions
300 (H) × 530 (W) × 300 (D) mm
Weight
23 kg
Request a quote

Product Details

1. Product Overview
The APSL Type 701 Series High Static Pressure Differential Pressure Fully Automatic Pressure Calibration System is a patented product of our company. The patents involved include: 1) Pressure calibration bench with built-in pressure module, patent number:201220302065.8, 2) Fully automatic pressure calibration bench for high static pressure differential pressure transmitters, patent number:201220302053.5, 3) Fully automatic pressure calibration bench based on HART communication, patent number:201220302062.4,4) Fully automatic pressure calibration bench, patent number:201220327679.1,5) Appearance of fully automatic pressure calibration bench, patent number:201230276258.6
This calibration bench is equipped with high-accuracy static pressure sensor and differential pressure sensor. When calibrating high static pressure differential pressure transmitter, set the static pressure set value and differential pressure range, press "Start" button, the system will automatically perform calibration: first open the bypass valve of differential pressure, automatically add static pressure, when static pressure reaches the set value, automatically according to the set number of calibration points, from the differential pressure range of0%~100% for calibration, automatically record data. If connected to a computer, it can automatically print the calibration report.

2. Performance Indicators
Working medium: air or transformer oil; no external high-pressure gas source is needed during calibration, and no need to exhaust air.
Static pressure module range: (0-20) MPa.
Differential pressure module range: (0-2000) kPa.
The internal pipeline diameter is relatively large, and a gas-liquid separator is installed, so it is not afraid of the calibrated meter carrying water or oil.
Measurement units: differential pressure: kPa. Static pressure: kPa or MPa.
Pressure display resolution: 5 digits.
Accuracy: pressure: class 0.02 or class 0.05, current: class 0.02 or class 0.05.
Stable accuracy: touch screen display ±5 digits.
Standard calibration points: range of0%、25%、50%、75%、100%, automatic return. Can be arbitrarily set when connected to a computer.
Custom calibration points: arbitrary.
Automatic pressure stabilization time at calibration points: (10~240) seconds, adjustable.
Total time for automatic calibration of transmitter (9-point stabilization): about 7 minutes.
Safety protection: overvoltage, overcurrent, pressure module fault, etc.
Current input: 1 channel,(0~22)mA。
Circuit power supply:DC24V/30mA, if mistakenly connected to a standard meter, it will not damage the standard meter.
Pressure interface:M20*1.5 internal thread quick coupling.
Current/voltage interface: Φ4 standard banana plug.
Drain port: inner diameter: Φ6mm。
Power supply:220V/50Hz。
Operating temperature: Operating temperature:-5℃~50℃。
Weight:23kg。
Dimensions: 300 (H) * 530 (W) * 300 (D) mm.
Automatic calibration software: optional.
External standard meter: optional.
3. Selection Instructions

Model Code

StaticPressureRange

PressureDifferenceRange

Accuracy code (Z

APSL-nmCYZ

code

Range

code

Measuring range

Code

DescriptionNotes

APSL:Series

0C

(0~50)kPa

KC

(0~2000)kPa

L

0.05Level True Color Touch Screen

n:Maximum Static Pressure Range

K

(0~20)MPa

GC

(0~1600)kPa

H

0.02Level True Color Touch Screen

m:Maximum differential pressure range

A

(0~10)MPa

AC

(0~1000)kPa

Model example:APSL-44CDL

Indicates static pressure range of(0~4)MPa, differential pressure range is (0~400)kPa, dual-channel, 0.05 class, touch screen controlled APSL type high static pressure differential pressure fully automatic pressure calibration bench

7

(0~7)MPa

7C

(0~700)kPa

6

(0~6)MPa

6C

(0~600)kPa

C:Range code

5

(0~5)MPa

4C

(0~400)kPa

Y:Channel code

4

(0~4)MPa

3C

(0~350)kPa

S:Single channel

3

(0~3.5)MPa

2C

(0~250)kPa

D:Dual channel

2

(0~2)MPa

1C

(0~100)kPa

Z:Accuracy code

1

(0~1)MPa

07

(0~70)kPa

Bottom maximum range

2000kPa

1600kPa

1000kPa

700kPa

600kPa

400kPa

350kPa

250kPa

200kPa

100kPa

70kPa

50kPa

Built-in differential pressure module1

2000kPa

1600kPa

1000kPa

700kPa

600kPa

400kPa

350kPa

250kPa

200kPa

100kPa

70kPa

50kPa

Built-in differential pressure module2

1000kPa

800kPa

500kPa

350kPa

300kPa

200kPa

160kPa

120kPa

100kPa

50kPa

35kPa

25kPa

Built-in differential pressure module3

500kPa

400kPa

250kPa

160kPa

150kPa

100kPa

80kPa

60kPa

50kPa

25kPa

16kPa

12kPa

Built-in differential pressure module4

250kPa

200kPa

100kPa

80kPa

70kPa

50kPa

40kPa

30kPa

25kPa

12kPa

8kPa

6kPa